Chapter 8 Operating Instructions
CLV490 Bar Code Scanner
8-10 © SICK AG · Division Auto Ident · Germany · All rights reserved 8009993/RA61/2007-07-20
Troubleshooting
8.5.2 Malfunctions in Reading mode: reading trigger errors
¾ Check whether the "Device Ready" LED lights up.
If the LED does not light, see Table 8-5, Page 8-9.
CLV490 in stand-alone configuration
Malfunction Possible cause Remedy
1. Trigger mode:
Sensor input (external sensor).
The CLV490 cannot be clocked:
the "Sensor" LED does not light up
the red scan line does not appear
Reading mode "Start/stop" or
"Tracking ":
Sensor not connected or connected to
incorrect switching input.
If sensor supplied by CLV490: sensor
not connected to ground potential.
Incorrect reading pulse source para-
meterized in the CLV490.
Polarity of switching input inverted
Incorrect debouncing of the switching
input parameterized in the CLV490.
Photoelectric switch not aligned with the
reflector (photoelectric proximity switch/
inductive sensor not dampened).
The object does not pass the reading
pulse sensor for trigger start with the
conveyor running.
Connect sensor to "Sensor" switching
input.
Check sensor wiring (see Fig. 5-5,
Page 5-22).
Measure output signal on sensor.
Insert jumper between pin 5 and pin 15
(CDM490: S 1 switch "SGND_GND")
In the "CLV-Setup" program: choose the
D
EVICE CONFIGURATION tab, READING TRIG-
GER SOURCE section: Sensor input select-
ed? (active high: reading pulse starts
when power applied, active low: reading
pulse starts when power removed)
See above.
In the "CLV-Setup" program: choose the
D
EVICE CONFIGURATION tab, click the
R
EADING TRIGGER PARAMETERS button,
DEBOUNCING SENSOR section:
Time controlled: selected value too high?
Track controlled: incremental encoder
connected to "IN 4" switching input?
Align sensor.
Check functioning
Install reading pulse sensor in front of the
CLV490 in conveying direction. See also
Chapter 4.4.2 Mounting the external rea-
ding pulse sensor, Page 4-11 .
Table 8-6: Troubleshooting: reading trigger errors in Reading mode (CLV490 in stand-alone configuration)